Type of hardness test | Rockwell and rockwell superficial |
Preliminary test force | 29.42N,98.07N |
Test force (rockwell superficial) | 147.1N,294.2N,441.3N |
Test force (rockwell) | 588.4N,980.7N,1471N |
Conformance | JIS B 7726, ISO 6508-2, ASTM E 18 |
Display | Digital |
Minimum reading | 0.1HR |
Preliminary test force (handling support) | Automatic steering wheel brake |
Preliminary test force setting | Dial switching |
Total test force setting | Dial switching |
Total test force control | Motor drive, automatic start |
Test force duration | 3~60s setting or manual operation |
Maximum specimen height | 180mm (100mm if cover is equipped) |
Maximum specimen depth | 165mm (from indenter axis to frame) |
Data output | Digimatic(SPC) , RS-232C |
Power | AC110-240V,1.2A |
Dimension (approx.) (W x D x H) | 235x516x780mm |
Weight (approx.) | 50kg |
*Specification and content are subjected to change without notice.