GT-7008-TR
Low Temperature Retraction Tester

The GT-7008-TR is specifically designed to assess the retraction characteristics of vulcanized rubber after being cooled to a low temperature. 
Specimens are first stretched at standard laboratory temperature and then cooled to a low temperature where no retraction occurs once the stretching force is removed. Subsequently, they are heated uniformly to determine the temperature at which a specified percentage of retraction takes place.

PRINCIPLE: Specimens are first stretched at standard laboratory temperature and then cooled to a low temperature where no retraction occurs once the stretching force is removed. Subsequently, they are heated uniformly to determine the temperature at which a specified percentage of retraction takes place.
WIDE MATERIAL APPLICATIONS: The GT-7008-TR is well suited for different types of vulcanized rubber used in various industries and applications.
HIGH STABILITY: Utilizes dedicated software to precisely maintain the stability and precision during stretching and retraction.
AUTOMATIC TEST PROCEDURE: The GT-7008-TR features automatic operation, enabling automatic testing procedures during stretching, cooling, and retraction. This minimizes human intervention and ensures testing accuracy.
HIGH PRECISION TEMPERATURE CONTROL: Equipped with high-precision temperature control system to ensure uniform heating rate during the retraction process.
COMPUTER-CONTROLLED TEMPERATURE RISE: Regulates temperature rise via dedicated software, ensuring unform rate of heating and reliable data during retraction.
STANDARD CONFORMITY: Ensures that products meet industry standards and satisfy quality expectations in the market.

Product Function Descriptionpdf download


Product Specification:ASTM D1329-2016、JIS-K 6261-2006 7.0、ISO 2921:2019、GB/T 7758-2020
Control system PLC+PC based software
Test station 3 or 6
Temp. range Ambient temp. to -70
Cooling system Compressor cooling system
Temp. accuracy ±0.3
Elongation range 350%
Displacement accuracy 0.01mm
Heating rate 1.0/min
Dimension (W×D×H) 105×85×235 cm